Used for only 1-2 hours, the doctor left the hospital and the system wasn’t in use!

MODULITH SLK intellect SYSTEM

Concept: Transportable shock wave system for x-ray and in-line ultrasound localization

Shock wave source positioning: motorized positioning of therapy head with selectable under- or over-table position (+45°lateral under- and -45° lateral over-table position)

In-line x-ray position in combination with urotable only.

Shock wave source:

  • Cylindrical electromagnetic shock wave coil with parabolic reflector
  • Shock wave penetration depth: 0 – 150 mm
  • Shock wave source diameter: 178 mm
  • Focus pressure: 6.4 – 92 MPa
  • Typical focus dimensions (Level 60): ∅ 6.3 mm x 62 mm
  • Energy flux density: 0.04 – 1.8 mJ/mm²
  • Shock wave frequency: 0.5, 1, 1.5, 2, 3, 4 SW/sec & ECG
  • Therapy head: Motorised positioning in X – Y – Z direction
  • Longitudinal movement: ±70 mm
  • Lateral movement: ±70 mm
  • Motorized height adjustment: 380 mm

SLK intelect operating panel:

  • Touch screen control of all system functions
  • Selection of shock wave source parameters
  • and shock wave release

Coupling Oil

Oil for proper coupling between the shockwave head and the patient film.

Hitachi Ultrasound Package 

Independent ultrasound system for inline stone localization and examination. The system features advanced ProSound technology, including a digital high-speed beamformer for multi-processing and subharmonic analysis (Broadband Harmonics) through quasi-compound vibration transmission. Additionally, it is equipped with a black-and-white and color Doppler system for electronic linear, electronic convex sector, and phased array sector transducers. The system offers imaging modes B/M/D as well as color flow and spectral Doppler (PW and CW, color flow and power flow, e-flow, TDI) with cine mode up to 12,800 frames in B-mode. Harmonic Echo Technology: Imaging with the second harmonic for artifact and noise suppression, available only for certain transducers (not available for inline transducers). Spatial Compound Imaging (SCI) for the suppression of artifacts dependent on beam direction, suppression of speckle noise, and easier diagnosis of images with higher contrast resolution. Silky Image Processing (SIP) simplifies differentiation (e.g., between tissue boundaries) by effectively removing artifacts. Optimal processing can be automatically configured for each clinical application. The system has connectors for three additional transducers. StorM-Link: Crosshair generation and imaging enhancement by changing the scan depth and using magnification. Inline localization concept and ultrasound-assisted patient positioning. 3.5 MHz inline transducer/convex transducer.

Scope of Delivery:

  • Hitachi F37 US Set including probe and inline unit
  • Hitachi DICOM Option SOP-F37-10 for F37-17

 Technical Data:

Scanning Method:

  • Electronic Convex Sector
  • Electronic Linear (steered scan and trapezoidal scan possible*)
  • Electronic Phased Array Sector *Dependent on the probe

Beamformer: Transmission: Semi-CPWG (Compound Pulse Wave Generator)

Reception:

  • Digital high-speed beamformer for multi-processing
  • 12-bit A/D converter (4096 gray levels)
  • Sampling frequency: Equivalent to 320 MHz
  • Delay accuracy: 1/64λ minimum for transmission and reception

Transmission/Reception of Harmonics:

  • Subharmonic analysis (Broadband Harmonics) through quasi-compound vibration transmission (dependent on the probe)
  • Reception frequency: 3.5-18.0 MHz* *18 MHz can be achieved under the following conditions:
    1. Probe: UST-5417
    2. Image frequency “High”
    3. Frequency information “Received”

Focusing:

  • Lateral direction
  • Transmission: Multistage transmission focus up to 3 out of 8 stages
  • Reception: Pixel focus
  • Slice direction
  • Acoustic lens

Beam Signal Processing: Dynamic apodization

System Dynamic Range: 180 dB

System Processing Channels: 199,872 channels

Frame Rate: Max. 745 frames/s or more* *Dependent on the probe and parameters

Abdominal Probe UST-9123 

Application: 2.5 – 6.0 MHz (60°/60 mm) Independent Diagnosis (B/W + Color) Harmonic Echo Imaging

Biopsy Needle Adapter MP-2473

Usage: Only for UST-9123

Application: Biopsy

Probe UST 91

EKG Mounting Kit 

Function: EKG-synchronized triggering of shock waves

Mounting: Portable

 

We can deliver worldwide and we can provide installation.

Should you have any questions, please don’t hesitate to ask.